Preventative Maintenance
General StatIM Preventative Maintenance:
Preventative Maintenance Schedule To ensure trouble-free performance, both the operator and the dealer must follow a preventative maintenance schedule.
NOTE: Please refer to your National, Regional, State or Safety laws for any additional reoccurring user testing that may be required.
Operator |
||
Daily |
Water Reservoir |
• Replace water as needed. • For opthalmic use, drain at the end of every workday, leave empty, and refill at the start of the next workday. |
| Waste Bottle |
• Empty the waste bottle every time you refill the water reservoir. • Fill the bottle with water, up to the MIN line marking. You may also add some chlorine-free disinfectant. |
|
Weekly |
Cassette |
• Wash the interior of the cassette with dishwashing soap or a mild detergent that does not contain chlorine. • Scrub the inside with a cleaning pad designed for use with Teflon™-coated surfaces. • After removing all traces of the detergent, treat interior surfaces of the cassette with the STAT-DRI™Plus drying agent to enhance the drying process. Order more STAT-DRI™ Plus from your dealer quoting 2OZPLUS, 8OZPLUST, or 32OZPLUS |
|
Biological and/or Air Filter |
• Check the filter for dirt and moisture. Replace if dirty. Call for service if wet. | |
|
Water Filter |
• Check the water reservoir filter every week and clean if necessary. Replace only if necessary. | |
Every 6 months |
Cassette Seal | • Replace every 500 cycles or six months (whichever is first), or whenever necessary. |
| Biological and/or Air Filter | • Replace every 500 cycles or six months (whichever is first). | |
Technician |
||
Once a year |
Cassette |
• Check the tray, lid and seal for damage. Replace if necessary. |
|
Biological Filter |
• Inspect the biological filter for moisture. | |
|
Solenoid Valve |
• Inspect the valve and clean if dirty. Replace the plunger if defective. | |
|
Pump |
• Clean the filters, replace if dirty. | |
|
Check Valve |
• Remove the exhaust tube from the back of the unit during the air drying phase. Check for air coming from the fitting. • Remove the air compressor tube from the check valve inlet while running a cycle. Make sure no steam is leaking from the valve. Replace if there are any leaks. |
|
|
Water Reservoir |
• Check the reservoir for dirt. Clean and rinse with steam process distilled water if necessary. | |
|
Calibration |
• Calibrate the unit. | |